2026, 4, 352-358
锥形束计算机断层扫描下研究不同固定装置对乳腺癌锁骨上淋巴引流区放疗摆位精度的影响
基金项目(Foundation):
邮箱(Email):
DOI: 10.3785/syzlzz.2026.049
移动端阅读
摘要:

目的 利用锥形束计算机断层扫描(cone-beam computed tomography, CBCT)探讨发泡胶固定和乳腺托架固定对包括锁骨上淋巴引流区(简称锁骨上靶区)的乳腺癌靶区放疗精度的影响。方法 回顾性选择2021年9月至2023年8月空军军医大学第一附属医院收治的74例术后行调强放射治疗的女性乳腺癌淋巴结转移患者,其中38例采用发泡胶固定(发泡胶组),36例采用乳腺托架固定(乳腺托架组)。所有患者每周接受1次CBCT扫描与定位CT计划图像进行匹配,并记录胸壁靶区和锁骨上靶区的摆位误差。比较2种固定方法的摆位误差。计算临床靶区扩至计划靶区(planning target volume, PTV)的外放边界(PTV margin, MPTV)。比较两组摆位时间和里克特五分制量表舒适满意度评分。采用Pearson相关系数分析2种固定方法摆位误差的相关性。结果 发泡胶组在胸壁靶区的左右方向(X轴)平移误差小于乳腺托架组[(1.6±1.5)mm vs(2.0±1.6)mm,P=0.012],锁骨上靶区的腹背方向(Z轴)平移误差[(1.4±1.3)mm vs(1.8±1.5)mm,P=0.046]和矢状位旋转误差[(0.5±0.5)°vs(0.7±0.5)°,P=0.006]均小于乳腺托架组。发泡胶组在胸壁靶区的X、Y(头脚方向)和Z轴的MPTV分别为5.01、5.99和5.47 mm,乳腺托架组分别为6.10、6.34和6.10 mm;发泡胶组锁骨上靶区X、Y和Z轴MPTV分别为3.69、3.86和4.28 mm,乳腺托架组分别为3.99、3.72和5.45 mm。发泡胶组分次摆位时间比乳腺托架组短[(3.4±1.1)min vs(5.5±3.1)min,P=0.007]。发泡胶组患者对固定装置的主观舒适满意度评分高于乳腺托架组[(27.5±1.2)分vs(25.4±1.2)分,P<0.01]。结论 乳腺癌淋巴结转移患者术后放疗中应用发泡胶固定在锁骨上靶区照射中的摆位精度比乳腺托架固定更高,MPTV更小,可提高工作效率和患者的治疗舒适度。

Abstract:

Objective To investigate the effects of foam cushion immobilization and breast board immobilization on positioning accuracy of breast cancer targets including the supraclavicular lymphatic drainage region(abbreviated as supraclavicular target volume) in radiotherapy by cone-beam computed tomography(CBCT). Methods A total of 74 female postoperative patients with lymph node-positive breast cancer receiving intensity-modulated radiation therapy who were admitted to the First Affiliated Hospital of Air Force Military Medical University from September 2021 to August 2023 were retrospectively enrolled. Among them, 38 patients receiving foam cushion immobilization were selected as the foam group, and 36 patients adopting breast board immobilization as the breast board group. All patients underwent weekly CBCT scanning registered with planning CT images. The setup errors of the chest wall target volume and supraclavicular target volume were compared between the two groups. The planning target volume margin(MPTV), expanded from clinical target volume(CTV) to planning target volume(PTV), was calculated. Single-fraction setup time and patient comfort and satisfaction scores based on the five-point Likert scale were compared between the two groups. Pearson correlation coefficient was applied to analyze the correlation of setup errors between the two immobilization methods. Results The translational setup error along the left-right(X-axis) direction of the chest wall target volume was significantly lower in the foam group than that in the breast board group [(1.6±1.5) mm vs(2.0±1.6) mm, P=0.012]. For the supraclavicular target volume, the anteroposterior(Z-axis) translational error [(1.4±1.3) mm vs(1.8±1.5) mm, P=0.046] and sagittal rotational error [(0.5±0.5)° vs(0.7±0.5)°, P=0.006] were significantly smaller in the foam group. The MPTV values on the X, Y(craniocaudal direction), and Z axes for the chest wall target volume were 5.01, 5.99, and 5.47 mm in the foam group, and 6.10, 6.34, and 6.10 mm in the breast board group, respectively, and those for the supraclavicular target volume were 3.69, 3.86, and 4.28 mm in the foam group, and 3.99, 3.72, and 5.45 mm in the breast board group, respectively. The single-fraction setup time was shorter in the foam group [(3.4±1.1) min vs(5.5±3.1) min, P=0.007]. The foam group had higher subjective comfort and satisfaction score [(27.5±1.2) points vs(25.4±1.2) points, P<0.01]. Conclusions For breast cancer patients with lymph node metastasis undergoing postoperative radiotherapy, foam cushion immobilization delivers superior setup accuracy and smaller MPTV during supraclavicular target irradiation compared with breast board immobilization. This technique can elevate clinical workflow efficiency and improve treatment-related comfort for patients.

参考文献

[1] Bray F, Laversanne M, Sung H, et al. Global cancer statistics 2022:GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries[J]. CA Cancer J Clin, 2024, 74(3):229-263.

[2] 高天,李照廷,肖敏.免疫检查点抑制剂在激素受体阳性乳腺癌新辅助治疗中的研究进展[J].实用肿瘤杂志, 2025, 40(1):1-6.

[3] Shao H, Sun YX, Na ZY, et al. Diagnostic value of applying preoperative breast ultrasound and clinicopathologic features to predict axillary lymph node burden in early invasive breast cancer:a study of 1247 patients[J]. BMC Cancer, 2024, 24(1):112.

[4] Tejada Ortigosa EM, Ollinger Casin I, Gaztelu Blanco I, et al. Ultra-hypofractionated 5-fraction radiation therapy for early breast cancer into whole breast and regional nodes:experience in a tertiary hospital[J]. Clin Transl Oncol,2025, 27(7):3135-3143.

[5] 廖恺,田允鸿,郑荣辉,等.早期鼻咽癌调强放疗设置颈前后部限量环减少头颈淋巴水肿的剂量学研究[J].实用医学杂志, 2024, 40(12):1659-1664.

[6] O'Connell AM, Karellas A, Vedantham S, et al. Newer technologies in breast cancer imaging:dedicated conebeam breast computed tomography[J]. Semin Ultrasound CT MRI, 2018, 39(1):106-113.

[7] Li B, Bai F, Yao XW, et al. Clinical value of styrofoam fixation in intracranial tumor radiotherapy[J]. Front Oncol, 2023, 13:1131006.

[8] Bai F, Zhang LH, Yao XW, et al. Neck muscle training improving the setup of reproducibility in patient with head and neck cancer receiving radiotherapy-a prospective randomized controlled study[J]. Radiat Oncol, 2025,20(1):35.

[9] Yang FT, Li JL, Li B, et al. Effectiveness of personalized open-face mask combined with styrofoam fixation in radiotherapy treatment of head and neck cancers:a prospective randomized controlled trial[J]. Radiat Oncol, 2025,20(1):34.

[10] Li J, Yang L, Yao XW, et al. A retrospective study on improving the accuracy of radiotherapy for patients with breast cancer with lymph node metastasis using Styrofoam[J]. Radiol Oncol, 2024, 58(1):124-132.

[11] Li XA, Tai A, Arthur DW, et al. Variability of target and normal structure delineation for breast cancer radiotherapy:an RTOG multi-institutional and multiobserver study[J]. Int J Radiat Oncol Biol Phys, 2009, 73(3):944-951.

[12] International Commission on Radiological Units and Measurements. ICRU Report 50:Prescribing, Recording, and Reporting Photon Beam Therapy[R]. Bethesda:ICRU,1993:22-26.

[13] International Commission on Radiological Units and Measurements. ICRU Report 62:Prescribing, Recording, and Reporting Photon Beam Therapy (Supplement to ICRU Report 50)[R]. Bethesda:ICRU, 1999:18-21.

[14] van Herk M, Remeijer P, Rasch C, et al. The probability of correct target dosage:dose-population histograms for deriving treatment margins in radiotherapy[J]. Int J Radiat Oncol, 2000, 47(4):1121-1135.

[15] Zhang L, Li B, Wang R, et al. Prospective randomized-controlled trial:efficacy of 37℃-preheated thermoplastic elastomer (TPE) bolus in postoperative radiotherapy for breast cancer patients[J]. Breast Cancer, 2025, 32:1375-1383.

[16] Mattke M, Rath D, Häfner MF, et al. Individual 3D-printed fixation masks for radiotherapy:first clinical experiences[J]. Int J Comput Assist Radiol Surg, 2021, 16(6):1043-1049.

[17] Vilotte F, Antoine M, Bobin M, et al. Post-prostatectomy image-guided radiotherapy:the invisible target concept[J]. Front Oncol, 2017, 7:34.

[18] Zhang HJ, Zhang C, Ge RG, et al. Evaluation of positional error during radiotherapy for breast cancer after modified radical mastectomy[J]. Biomed Res, 2017, 28(17):7526-7533.

[19] Lv R, Yang GY, Huang YZ, et al. Dosimetric effects of supine immobilization devices on the skin in intensity-modulated radiation therapy for breast cancer:a retrospective study[J]. BMC Cancer, 2021, 21(1):384.

[20] Dinu A, Flonta T, Virag V, et al. Breast irradiation using personalised thermoplastic mask for immobilisation:Pros and cons[J]. Eur J Cancer, 2017, 72:S28.

[21] 周常锋,房建南,黄晓波,等.聚氨酯发泡胶与负压真空垫在乳腺癌保乳根治术后IMRT体位固定精度初步研究[J].中华放射肿瘤学杂志, 2019, 28(10):776-779.

[22] 房建南,马玉家,石俊田,等.发泡胶与乳腺托架在乳腺癌保乳术后IMRT中固定精度比较[J].中华放射肿瘤学杂志, 2019, 28(5):369-372.

[23] 张彦新,郇福奎,朱耿强,等.乳腺癌术后调强放疗颈胸一体固定的摆位误差分析[J].中华放射肿瘤学杂志,2021, 30(8):835-840.

[24] Shen KL, Xiong J, Wang ZG, et al. Design of a new breast vacuum bag to reduce the global and local setup errors and to reduce PTV margin in post-mastectomy radiation therapy[J]. J Radiat Res, 2020, 61(6):985-992.

[25] Taylor C, Correa C, Duane FK, et al. Estimating the risks of breast cancer radiotherapy:evidence from modern radiation doses to the lungs and heart and from previous randomized trials[J]. J Clin Oncol, 2017, 35(15):1641-1649.

[26] Vasiljevic D, Arnold C, Neuman D, et al. Occurrence of pneumonitis following radiotherapy of breast cancer-a prospective study[J]. Strahlenther Und Onkol, 2018,194(6):520-532.

[27] Duffton A, Li W, Forde E. The pivotal role of the therapeutic radiographer/radiation therapist in image-guided radiotherapy research and development[J]. Clin Oncol,2020, 32(12):852-860.

[28] Zhou JD, Li S, Ye CW, et al. Analysis of local setup errors of sub-regions in cone-beam CT-guided post-mastectomy radiation therapy[J]. J Radiat Res, 2020, 61(3):457-463.

[29] Dong F, Weng X, Deng X, et al. Clinical utility of a new immobilization method in image-guided intensity-modulated radiotherapy for breast cancer patients after radical mastectomy[J]. J Xray Sci Technol, 2022, 30(4):641-655.

[30] Lin CG, Xu SK, Yao WY, et al. Comparison of set up accuracy among three common immobilisation systems for intensity modulated radiotherapy of nasopharyngeal carcinoma patients[J]. J Med Radiat Sci, 2017, 64(2):106-113.

[31] 姚文燕,刘镖水,方键蓝,等.乳腺托架加发泡胶提升乳腺癌放疗体位固定精度的研究[J].中华放射肿瘤学杂志, 2022, 31(10):916-921.

[32] Svestad JG, Heydari M, Mikalsen SG, et al. Surface-guided positioning eliminates the need for skin markers in radiotherapy of right sided breast cancer:a single center randomized crossover trial[J]. Radiother Oncol, 2022,177:46-52.

[33] Harron EC, McCallum HM, Lambert EL, et al. Dosimetric effects of setup uncertainties on breast treatment delivery[J]. Med Dosim, 2008, 33(4):293-298.

[34] 田素青,孙海涛,赵田地,等. 6D治疗床辅助影像引导下放射治疗头颈部肿瘤摆位误差分析[J].北京大学学报(医学版), 2024, 56(6):1097-1100.

[35] 冯鑫,郇福奎,陈欢,等.光学体表监测系统在乳腺癌改良根治术后放疗中的应用研究[J].实用肿瘤杂志,2023, 38(1):59-65.

基本信息:

DOI:10.3785/syzlzz.2026.049

引用信息:

[1]周艳,姚晓伟,张丽华,等.锥形束计算机断层扫描下研究不同固定装置对乳腺癌锁骨上淋巴引流区放疗摆位精度的影响[J],2026,41(4):352-358.DOI:10.3785/syzlzz.2026.049.

检 索 高级检索